Unit 6: Practical Skills in Physics A2 II
- Syllabus
- 2021
- Section
- —
- Level
- A2

Choose apparatus by working backwards from the variable, its expected size and the precision needed to test the proposed relationship. A named instrument is incomplete unless its usable range and resolution fit the measurement.
| Detail | Planning question |
|---|---|
| range | Can the instrument measure the smallest and largest expected values? |
| resolution | Is the smallest scale change small enough compared with the measured value? |
| apparatus dimensions | Do length, diameter, volume or separation permit the intended range and method? |
| measurement access | Can the relevant points be aligned, contacted or detected without changing the system? |
percentage uncertainty≈measured valueabsolute uncertainty×100%
For a diameter near 20mm, vernier calipers with 0.1mm resolution give an uncertainty small compared with the diameter and are more suitable than a millimetre ruler. A micrometer may be preferable for a thickness near 1mm.
Highest resolution is not automatically most appropriate: the instrument must also have enough range and suit the object. State numerical range, resolution or relevant dimensions when the context makes them important.
Calibration checks whether an instrument's indication corresponds to known input values. Before measurements, check the zero where appropriate; a non-zero reading with zero input reveals a zero error.
| Step | Purpose |
|---|---|
| apply zero input and record the indication | detect a zero offset |
| compare one or more readings with known standards | test the scale response across the intended range |
| adjust the instrument or record a correction | prevent the offset entering every result |
| repeat the check during a long investigation if drift is plausible | confirm calibration remains valid |
If an instrument reads +0.04mm when it should read zero and cannot be reset, subtract 0.04mm from every raw reading. Record both the check and the correction in the plan.
Calibration chiefly addresses systematic error: an offset can shift all readings in the same direction, so taking a mean of repeated uncorrected readings does not remove it.
A zero check is one form of calibration, not the whole idea. A zero reading does not prove that the instrument is accurate throughout its range; comparison with known values may also be needed.
For every measured variable, state the instrument, the exact technique and how its reading becomes the required quantity. The description should let another person reproduce the measurement without guessing reference points, timing events or conversions.
| Variable | Executable measurement detail |
|---|---|
| diameter or thickness | use vernier calipers or a micrometer; close correctly, read at eye level and sample relevant positions/orientations |
| oscillation period | time several complete oscillations between repeated passages of a fixed marker, then divide by the number |
| count rate | count for a stated long interval, measure background over the same interval and subtract background rate |
| potential difference | connect a voltmeter in parallel across the named component and select a suitable range |
State how the independent variable will be changed and measured for at least a useful spread of values. Define the dependent-variable reading at each setting and include any necessary derived quantity.
Use techniques that reduce avoidable reading error: fixed fiducial markers for timing, perpendicular viewing to limit parallax, and unambiguous centre-to-centre or end-point definitions for distance.
Writing only 'measure x' or naming an instrument is not a method. Do not confuse measurement technique with later repetition: first make each individual reading valid, then decide whether and how to repeat it.
A fair test changes the independent variable, measures the dependent variable and holds other variables that could affect the result constant. For every control variable, state both how it is controlled and why its variation would matter.
| Planning move | Example for light intensity on a solar cell |
|---|---|
| identify a competing influence | source-to-cell distance also changes intensity |
| give an operational control | clamp source and cell at a fixed measured separation |
| explain the consequence | changing distance would alter the voltmeter reading independently of the chosen variable |
| monitor another relevant condition | exclude or keep background light constant |
Choose controls from the physics of the system rather than listing generic conditions. Temperature, alignment, initial displacement, source-detector distance or component values matter only when they can change the dependent variable in that experiment.
A useful sentence pattern is: 'Keep ___ constant by ___ because ___ would otherwise change ___.'. This makes the variable, method and causal reason explicit.
Do not call the independent variable a control variable, and do not write 'keep everything else the same'. A control must be relevant, measurable or enforceable, and tied to the validity of the comparison.
Repeat a reading under the same stated conditions when random variation is plausible and the measurement can be reproduced. Repeats reveal scatter, help identify an anomalous result and allow a representative mean.
| Situation | Repeat strategy |
|---|---|
| manual timing or a fluctuating sensor reading | repeat at each setting and calculate a mean |
| period of a rapid oscillation | time many cycles per run, then repeat runs |
| variation across an object, such as thickness | measure at several relevant positions before finding a mean |
| counting random events | use a sufficiently long interval and repeat if time permits |
xˉ=nx1+x2+⋯+xn
A planning answer should say whether repeats are appropriate, how many or at which settings, what will be averaged, and how an anomalous value will be checked. The choice should match the source of variation.
Repeating does not correct a zero error, poor calibration or a consistently biased technique. It reduces uncertainty associated with random scatter only when the conditions genuinely remain the same.
Identify each hazard created by the proposed apparatus or procedure, state the harm it could cause, then specify a control that reduces likelihood or exposure without preventing the measurement.
| Hazard | Possible harm | Proportionate control |
|---|---|---|
| charged capacitor | electrical discharge | keep within working p.d., observe polarity and discharge before handling |
| ionising-radiation source | tissue exposure | maximise distance, minimise time and use tongs; point away from people |
| laser beam | eye damage | never view directly; keep beam below eye level and use a stop |
| hot component | burn | switch off, allow cooling and handle with suitable protection |
| falling or moving mass | impact | secure clamps and keep the fall zone clear |
Prefer removing or enclosing the hazard and limiting exposure before relying on personal protection. Include only controls relevant to the actual setup and explain what each control manages.
Safety is part of the executable method: specify when equipment is isolated, discharged, cooled, shielded or repositioned. If a hazard remains significant, change the apparatus or method.
Generic phrases such as 'take care', 'wear goggles' or 'do a risk assessment' earn little scientific value without a named hazard and matched control. Do not invent hazards that the proposed experiment does not contain.
State how the measurements will test the proposed relationship. This determines which raw quantities, repeats, range and derived columns must be collected; analysis is therefore part of planning, not an afterthought.
| Stage | Planned use of data |
|---|---|
| process | calculate means and required derived quantities with units |
| display | plot the dependent or transformed variable on the vertical axis against the independent or transformed variable on the horizontal axis |
| test | compare the graph's shape, straightness, gradient or intercept with the prediction |
| conclude | state whether the data support the relationship within scatter and uncertainty |
Derive the graph from the proposed model. For constant acceleration from rest, x=frac12at2, so plotting t2 against x should give a straight line through the origin with gradient 2/a. For T2=C(2l−h), plotting T2 against h should be linear with the predicted sign and intercept relationship.
Plan enough well-spaced values to reveal the trend rather than merely comparing two endpoints. State how the best-fit line or curve will be used; do not promise a conclusion unsupported by the chosen plot.
'Plot a graph' is incomplete unless both axes and the expected test are named. Correlation or approximate straightness supports a proposed model but does not by itself prove that no other physical explanation is possible.
Evaluate an experimental method by linking a specific weakness to a practical modification and then explaining how the change improves the measurement. The causal link matters more than naming sophisticated apparatus.
| Observed weakness | Targeted modification | Why it helps |
|---|---|---|
| cylinder diameter is difficult to align | place set squares against opposite sides and measure their separation | defines tangents and reduces alignment error |
| timing point changes between oscillations | use a fixed marker at the equilibrium position | gives the same timing event each pass |
| human reaction time limits a short interval | use a light gate or motion sensor with a data logger | automates detection and provides closely spaced readings |
| count rate includes background | measure background under the same conditions and subtract it | removes the background contribution from the measured rate |
Discuss limitations as well as benefits. A light gate removes human reaction time, for example, but its trigger position or the part of an object that breaks the beam may still create uncertainty. A video helps only if its frame rate and scale suit the event.
Quantify the advantage when possible: a finer resolution, faster sampling, simultaneous readings or a longer timing interval should reduce a named uncertainty or prevent a named bias.
Do not claim vaguely that a data logger 'makes results accurate'. State which measurement changes and why. Repeats reduce random scatter; they do not repair a systematic offset or an unsuitable technique.
Judge the number of readings from the intended conclusion. There must be enough independent-variable settings to reveal the shape or key feature of the relationship, and enough repeated readings where random scatter affects each setting.
| Count to evaluate | What too few readings prevent | Appropriate improvement |
|---|---|---|
| distinct settings of the independent variable | reliable trend, curve or comparison | add well-spaced settings across the region of interest |
| settings near a peak or turning point | accurate location of that feature | add more closely spaced readings around the feature |
| repeats at one setting | estimate of scatter and reliable mean | repeat under unchanged conditions and calculate a mean |
If a resonance curve has only one or two readings between 1.4 and 1.6Hz, the peak frequency is poorly located even if many points exist far from the peak. More readings should be concentrated around the maximum.
Use the actual table or graph: count the distinct values, inspect their spacing and note whether repeats are shown. Explain what additional readings would make the conclusion more secure.
A larger total count is not automatically better. Ten repeated readings at one setting cannot replace missing independent-variable values, and many distant points cannot locate a narrow peak.
The range is the interval from the smallest to the largest value of the independent variable. Evaluate whether it is wide enough to expose the predicted change while remaining within safe, measurable and physically valid conditions.
| Range question | Why it matters |
|---|---|
| Is the dependent-variable change large compared with reading scatter? | a narrow range may hide the trend |
| Does the interval include the predicted peak, threshold or curvature? | a missing feature cannot be located or tested |
| Are values spread across the whole interval? | clustered values leave parts of the relationship unsupported |
| Do instruments remain within range and the model's conditions? | extreme values may saturate equipment or invalidate assumptions |
A spring investigation using extensions only from 1.0 to 1.5cm may show too little force change to judge proportionality. Extending the interval can clarify the trend, but only while the spring remains below its elastic limit.
State a justified new minimum or maximum when the context permits, not merely 'use a wider range'. The proposed extension should target the weak evidence without changing the physical question.
Range and number of readings are separate. A wide interval with only two points has poor coverage; many closely spaced points over a tiny interval still have a narrow range.
A results table must not imply more precision than the apparatus and input data support. Check raw readings by instrument resolution and calculated values by the precision of the measurements used.
| Result type | Appropriate recording rule |
|---|---|
| analogue scale reading | record to the scale resolution, with consistent decimal places in the column |
| digital reading | record all displayed digits unless the display is known to fluctuate meaninglessly |
| repeated readings of one quantity | use a consistent decimal place determined by the instrument |
| calculated result | round to a sensible number of significant figures, normally no more than the least precise input justifies |
If measured values 3.4 and 1.26 are multiplied, the calculator gives 4.284, but 3.4 has only two significant figures, so reporting 4.3 is appropriate. Writing 4.28400 would invent precision.
Scan each column for one entry with extra or missing digits, then use the heading, instrument detail and neighbouring entries to decide the correction. Keep leading zeros non-significant: 0.00450 has three significant figures.
Significant figures and decimal places are not interchangeable. Raw readings from the same instrument usually need consistent decimal places; derived results are rounded according to justified significant figures.
Identify a unit by tracing what quantity was measured or calculated. A table heading should normally separate the quantity symbol from its unit, and every derived or transformed column needs the corresponding derived unit.
| Quantity or operation | Correct unit consequence |
|---|---|
| length x measured in metres | x/m |
| area A=x2 | m2, not m |
| reciprocal time 1/t | s−1 |
| squared time t2 | s2 |
| gradient Δy/Δx | unit of y divided by unit of x |
Convert prefixes consistently before calculation: 1mm=10−3m, so 1mm2=10−6m2. Squaring a value also squares its conversion factor.
To amend an incorrect heading, first identify the underlying quantity and transformation, derive the unit algebraically, then rewrite the unit without changing the numerical data unless a unit conversion is also required.
A unit is not chosen from the appearance of the numbers. Do not leave a transformed column with the original unit, and do not square only the unit label while forgetting the numerical conversion.
An inconsistent reading is one that does not follow the pattern of the other results beyond their normal scatter. Locate it in the table or as a point displaced from the graph trend, then investigate before deciding how it should be treated.
| Check | Possible finding |
|---|---|
| compare graph point with the original table | plotting or transcription error |
| repeat the calculation and unit conversion | arithmetic or prefix error |
| inspect apparatus notes and conditions | changed control variable, poor contact or misalignment |
| repeat the measurement at the same setting | original reading was anomalous, or the unusual result is reproducible |
| take nearby settings | genuine local feature rather than an isolated error |
If a repeat under controlled conditions agrees with the surrounding trend and not the original value, record the original as anomalous and justify excluding it from a mean or fit. If the unusual value repeats, retain it and reconsider the assumed relationship or method.
Judge inconsistency relative to scatter and measurement precision. A point slightly off a best-fit line is not automatically anomalous; experimental data are not expected to lie exactly on a mathematical line.
Never delete a point solely because it spoils the expected pattern. Evidence for a recording, calculation or measurement problem—or a failed repeat—is needed before exclusion is defensible.
Report a calculated result only to the precision justified by the measured inputs. Keep extra calculator digits during the working, then round once at the end so intermediate rounding does not distort the final value.
| Calculation | Practical reporting rule |
|---|---|
| multiplication or division | use no more significant figures than the least precise measured input |
| addition or subtraction | use no more decimal places than the least precise term justifies |
| exact count or defined conversion | it does not limit significant figures |
| uncertainty stated with a result | round the value to the same decimal place as the uncertainty |
For v=1.25m/0.42s, the calculator gives 2.976…ms−1. The time has two significant figures, so report v=3.0ms−1, not every displayed digit.
Leading zeros are placeholders, while trailing zeros after a decimal can show precision: 0.00450 has three significant figures. Scientific notation makes the intended precision unambiguous.
Do not round every intermediate line. Also do not force a universal number such as three significant figures; use the actual input precision and the context of the measurement.
Choose axes that test the intended relationship, label each quantity and unit, and use a simple scale that spreads the data across most of the available grid. Plot points accurately before drawing a best-fit line or curve.
| Feature | Good practice |
|---|---|
| axes | independent or chosen transformed variable horizontally; dependent variable vertically |
| labels | quantity symbol and unit, such as T/s |
| scale | uniform, easy intervals; need not begin at zero unless zero is physically or analytically important |
| points | small crosses or clear points positioned within plotting precision |
| fit | one balanced best-fit line or smooth curve; do not join dot-to-dot |
For a power law y=kxn, form dimensionless processed variables Y=log(y/y0) and X=log(x/x0) using stated reference units. Then Y=nX+C, where C depends on k and the chosen reference units. Plotting Y against X on linear graph paper can reveal a straight line.
A log-log graph labelled log(T/s) vertically and log(M/kg) horizontally has dimensionless plotted values. Use consistent decimal places for the processed logarithms.
A visually steep graph is not automatically more sensitive or more convincing: axis scale changes appearance. Judge the fit, scatter and derived gradient using the labelled numerical scales.
Carry units through substitutions, transformations, graph labels, gradients and final constants. Unit algebra is an independent check that the calculation and chosen relationship are physically consistent.
| Analysis step | Unit result |
|---|---|
| v=s/t | ms−1 |
| t2 | s2 |
| 1/x for x in metres | m−1 |
| gradient of y against x | unit of y divided by unit of x |
| logarithm | use a dimensionless ratio such as log(T/s) |
Convert prefixes before combining quantities. If d=2.4mm=2.4×10−3m, then d2=5.76×10−6m2; both the number and the unit conversion are squared.
When a gradient represents a physical constant, derive its unit from the plotted axes and compare it with the unit expected from the model. A mismatch often exposes a reversed axis, missed power or unconverted prefix.
Writing an SI-looking unit at the end cannot repair inconsistent substitutions. Do not take a logarithm of a dimensional quantity without expressing it as a ratio to the stated unit.
A trend comment states what the data show: direction, form, important region and scatter. Separate this observation from a physical explanation or causal claim unless the evidence and controlled design support that conclusion.
| Data feature | Precise description |
|---|---|
| straight line through the origin | variables may be directly proportional within uncertainty |
| straight line with non-zero intercept | linear relationship, but not direct proportionality |
| decreasing curve | one variable decreases non-linearly as the other increases |
| constant ratio y/x | supports y∝x |
| isolated displaced point | possible anomalous result requiring a check |
Transform the proposed relationship when useful. To test T∝D, check whether T/D is approximately constant or whether T2 plotted against D is linear through the origin.
Use data to support the comment: quote a range, ratio or change and acknowledge scatter. 'Increases' is weaker than 'approximately linear with a small positive intercept'.
A positive correlation does not by itself prove that the horizontal-axis variable causes the change. Nor does a roughly straight line prove proportionality unless the intercept is consistent with zero.
Rearrange the proposed model into Y=mX+c, choose plotted quantities Y and X, then interpret the gradient m and intercept c in terms of the required relationship or constant.
m=ΔXΔY=X2−X1Y2−Y1
Choose two well-separated points on the best-fit line, not necessarily measured data points. Draw a large triangle spanning at least about half the plotted line, read coordinates carefully and keep the subtraction signs. A large triangle reduces the percentage effect of coordinate-reading uncertainty.
| Linear plot | Meaning |
|---|---|
| y against x | gradient gives the coefficient of x; intercept gives the constant term |
| y against x2 | linearity supports y∝x2 when intercept is consistent with zero |
| Y=log(y/y0) against X=log(x/x0) for y=kxn | gradient =n; intercept is the constant set by k and the reference units |
Give the gradient an appropriate sign, significant figures and unit derived from vertical-axis unit divided by horizontal-axis unit. A log-log gradient is dimensionless.
Do not calculate a gradient from two neighbouring raw points or from the edges of the paper unless they lie on the best-fit line. The axes determine what the gradient physically represents.
| Term | Meaning | Evidence |
|---|---|---|
| precision | repeated values are close to one another | small spread or range |
| accuracy | a result is close to the accepted or true value | accepted value lies within the uncertainty interval, or percentage difference is suitably small |
| sensitivity | instrument output changes substantially for a small change in input | large change in output per unit input; steep calibration response |
Two sets can have the same mean but different ranges: the set with the smaller range is more precise. Without an accepted value, their accuracy cannot be decided from agreement alone.
Greater sensitivity can make small input changes easier to distinguish, but sensitivity is not the same as resolution, precision or accuracy. An instrument can respond strongly yet retain a calibration offset.
Use uncertainty when judging accuracy: if g=10.0±0.6ms−2, the accepted 9.81ms−2 lies in the interval, so the result is consistent with it.
Precise results can all be inaccurate because of systematic error, and one accurate-looking value does not prove a precise method. Use each term only for the property it describes.
A realistic error-reduction proposal names the error source, changes one practical feature and explains which uncertainty or bias becomes smaller. Match the modification to random or systematic behaviour.
| Error source | Realistic modification | Effect |
|---|---|---|
| zero offset | check zero and reset or apply a correction | reduces systematic bias |
| reaction time in a short timing | time many cycles or use automatic sensing | reduces percentage timing uncertainty |
| parallax in a scale reading | view normally with a fixed pointer or mirror alignment | reduces reading bias |
| variation across an object | measure at several positions and average | reduces effect of random spatial variation |
| fluctuating value | repeat under the same conditions and calculate a mean | reduces effect of random scatter |
Prefer a change whose effect can be predicted. Timing ten cycles makes the measured interval about ten times larger while similar start-stop uncertainty remains, so percentage uncertainty in the period decreases.
The proposal must suit the existing experiment, available range and safety constraints. Explain any new limitation introduced by additional apparatus, such as sensor alignment or sampling rate.
Repeats do not remove a systematic offset, and higher resolution does not correct poor calibration. Avoid the impossible promise to 'eliminate all error'; aim to reduce a named contribution.
An experiment is improved when its method can support the intended conclusion more securely. Diagnose the weakest link—validity, data coverage, control, measurement quality or reproducibility—then propose a feasible change and explain the evidential gain.
| Weakness | Improvement | Stronger evidence |
|---|---|---|
| too few independent-variable settings | add well-spaced values and extra points near important features | trend or peak is defined more reliably |
| narrow measurement range | extend within apparatus and model limits | predicted variation is easier to distinguish from scatter |
| uncontrolled competing variable | measure and hold it constant | change in the dependent variable is more attributable to the intended cause |
| result cannot be independently checked | repeat the whole method or compare with a second valid technique | reproducibility or method dependence becomes visible |
State how the modification changes the decision made from the data. If it adds time, complexity or a new uncertainty, weigh that cost against the expected improvement.
Use details from the actual setup and results. A good proposal names the variable, apparatus setting or data region to change rather than giving a generic list of laboratory virtues.
More apparatus or more data are not automatically improvements. A change is useful only if it addresses the observed weakness while keeping the physical test and safety conditions valid.
Measurement uncertainty expresses a justified interval around a reported value. Write x±Δx, where Δx is the absolute uncertainty in the same unit as x; it reflects resolution, scatter and method limitations rather than a known error from the true value.
fractional uncertainty=xΔx,percentage uncertainty=xΔx×100%
Discuss the dominant source and direction when relevant: limited resolution creates a reading interval, repeats reveal random spread, and an uncorrected zero error could shift all values. State whether a proposed change reduces the absolute or percentage contribution.
Compare an accepted value or another result with the uncertainty interval. If E=14.3±0.7GPa, values from 13.6 to 15.0GPa are consistent with the measurement; a candidate value outside that interval is not.
Uncertainty is not automatically the difference from an accepted value, and it is not evidence that the measured value is definitely wrong. It describes the measurement's supported interval under the stated method.
Propagate uncertainties by following how measured quantities are combined. The standard school-level rules give a maximum estimated uncertainty for independent contributions.
| Relationship | Combine uncertainties |
|---|---|
| Q=a±b | add absolute uncertainties: ΔQ=Δa+Δb |
| Q=ab or Q=a/b | add percentage uncertainties |
| Q=an | multiply the percentage uncertainty in a by ∣n∣ |
| constants such as 2 or π | exact constants add no measurement uncertainty |
Q=crapbq⟹%UQ=∣p∣%Ua+∣q∣%Ub+∣r∣%Uc
For R=V/I, if V has 2.0% uncertainty and I has 3.0%, then R has approximately 5.0% uncertainty. If A=πd2/4 and d has 1.5%, then A has 3.0% uncertainty.
After finding the combined percentage uncertainty, convert to an absolute uncertainty when an interval is needed: ΔQ=(%UQ/100)Q. Round the uncertainty sensibly and match the value's decimal place.
Do not add raw absolute uncertainties for multiplication, and do not square the percentage uncertainty when a variable is squared—the exponent multiplies it.
For one reading, use half the instrument resolution as the absolute uncertainty. If a scale's smallest division is 1mm, a single length reading has Δx=0.5mm under this syllabus convention.
%U=reading21(resolution)×100%
For repeated readings, calculate the mean and use half the range as the absolute uncertainty. The range is maximum minus minimum.
xˉ=n∑xi,Δx=2xmax−xmin,%U=xˉΔx×100%
Repeated times for five cycles are 4.84, 4.92, 4.88 and 4.96s. Their mean is 4.90s and half range is (4.96−4.84)/2=0.06s. For one period, divide both by five: T=0.980±0.012s, with the same percentage uncertainty, about 1.2%.
Do not use the full resolution for a single reading or the full range for repeats in this specification. When a total for several cycles is divided, divide its absolute uncertainty by the same number.